Serial dilution cheat sheet
Everything you need to plan an N-step dilution ladder from a stock, in one page.
The identities
Starting from a stock , after N uniform D-fold dilutions:
To reach a target concentration in exactly N steps, solve for the per-step factor:
Each step transfers of solution from the previous tube into fresh diluent, so the total volume per tube stays fixed at :
Stock consumed total = . Diluent used per step = .
Common ladders
| Ladder | D per step | Steps to 10⁶-fold | Transfer (per 100 µL step) | Notes |
|---|---|---|---|---|
| 1:2 | 2 | 20 | 50 µL into 50 µL diluent | Antibody titers, IC₅₀ dose-response |
| 1:3 | 3 | 12.6 (round up) | 33 µL into 67 µL diluent | Uneven; use for dose response with 3-fold steps |
| 1:5 | 5 | 8.6 | 20 µL into 80 µL diluent | Compromise between resolution and step count |
| 1:10 | 10 | 6 | 10 µL into 90 µL diluent | Default choice — clean powers of 10 |
| 1:100 | 100 | 3 | 1 µL into 99 µL diluent | 1 µL is at the P2 floor — prefer 1:10 |
The 1:10 ladder is the workhorse: powers of 10 are easy to reason about, transfers stay comfortably in P20 range, and 6 steps hits the 10⁶-fold dilutions common in microbiology (CFU counts) and PCR (standard curves).
Pipette accuracy floors
Every pipette has a low-end below which the CV climbs. Below these values, spread across replicate transfers is ≥ 5 % and you should split the step further.
| Pipette | Range | Accurate down to | Do not use below |
|---|---|---|---|
| P2 | 0.1 – 2 µL | 0.2 µL | 0.1 µL |
| P10 (if available) | 0.5 – 10 µL | 1 µL | 0.5 µL |
| P20 | 2 – 20 µL | 2 µL | 1 µL |
| P100 | 10 – 100 µL | 10 µL | 5 µL |
| P200 | 20 – 200 µL | 20 µL | 10 µL |
| P1000 | 100 – 1000 µL | 100 µL | 50 µL |
When to serial dilute
Any time a one-step would call for a stock volume below the smallest pipette’s accurate floor. Rough test:
- ≤ 1 µL → serial-dilute.
- Total dilution factor 100× → serial-dilute in ≥ 2 steps.
- Total dilution factor 10⁵× → serial-dilute in ≥ 3 steps.
Choosing N
Two constraints trade off:
- Small N → fewer transfers → less pipetting error, but bigger D per step and possibly a below-floor transfer.
- Large N → each step is a comfortable D, but every transfer adds ~1 % CV.
Heuristic. Pick the smallest N such that every step’s transfer volume is ≥ 2× the pipette floor. For most 10⁶ dilutions, that lands on N = 6 with D = 10.
Diluent volume trick
For a step volume of and a dilution factor of D:
For a 1:10 ladder at 100 µL steps: µL. For a 1:2 ladder at 100 µL steps: µL.
Common mistakes
- Not vortexing / not changing tips between steps. Carryover of stock down the ladder inflates every subsequent concentration. Change tips.
- Pipetting below the pipette floor. A 0.5 µL transfer with a P20 has ≥ 10 % CV. Split into two steps instead.
- Mismatched step volumes. If isn’t constant across the ladder, D per step isn’t either, and the analytical formula silently breaks. Every step must use the same volume, or you must track each / pair separately.
- Not accounting for compound stickiness. Peptides, hydrophobic small molecules, and DMSO stocks can lose 5–20 % to plastic surfaces per step. Consider low-bind tubes or a carrier protein.
Sources
- Rainin RC-500 / RC-1000 pipette specifications — accuracy tables
- Gilson Pipetman NEO product spec sheet — CV at low-end volumes
- Harris, Quantitative Chemical Analysis, 10th ed. (2020), §2-2 — analytical precision
Last updated: .